FDA Adverse Event Malfunction Summary report: N

12.4 CM HEAVY DUTY LONG ATTACHMENT

MDR report key: 4130647 · Received September 30, 2014

Report

Report Number
1045834-2014-12966
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
September 19, 2014
Report Date
September 19, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR SERVICE. DURING PRE-REPAIR ASSESSMENT, IT WAS DETERMINED THAT THE DEVICE DID NOT MEET MANUFACTURING SPECIFICATIONS. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. IT WAS DETERMINED THAT THIS IS MOST LIKELY DUE TO EXCESSIVE SIDE LOADING, WHICH CAUSED THE BEARINGS TO GO BAD AND THE DEVICE TO OVERHEAT. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO MISUSE, ABUSE AND/OR USER ERROR. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SERVICE AND REPAIR/PRE-TESTING, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE HAD HIGH TEMPERATURE. THIS EVENT IS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
608187 12.4 CM HEAVY DUTY LONG ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1