PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-96173
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Date of Event
- July 22, 2014
- Report Date
- July 22, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP ALARMED MOTOR ERROR DURING THE OCCLUSION TEST AND ALARMED PRIME DURING PRIME TEST AT 4LBS DUE TO FAULTY FORCE SENSOR. MOTOR PASSED MOTOR TEST. INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNER, RESERVOIR TUBE LIP AND MINOR SCRATCHES ON DISPLAY WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE CUSTOMER WAS CHANGING RESERVOIR AND GOT COMPROMISED FORCE SENSOR SYSTEM ALARM. CUSTOMER BLOOD GLUCOSE WAS 210 MG/DL. TROUBLESHOOTING WAS DONE. DURING TROUBLESHOOTING CUSTOMER HAD MOTOR ERROR ALARM DURING BOLUS DELIVERY. ADVISED CUSTOMER THAT INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 609384 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR |