FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4130616 · Received September 30, 2014

Report

Report Number
3004209178-2014-96173
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
July 22, 2014
Report Date
July 22, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP ALARMED MOTOR ERROR DURING THE OCCLUSION TEST AND ALARMED PRIME DURING PRIME TEST AT 4LBS DUE TO FAULTY FORCE SENSOR. MOTOR PASSED MOTOR TEST. INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNER, RESERVOIR TUBE LIP AND MINOR SCRATCHES ON DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS CHANGING RESERVOIR AND GOT COMPROMISED FORCE SENSOR SYSTEM ALARM. CUSTOMER BLOOD GLUCOSE WAS 210 MG/DL. TROUBLESHOOTING WAS DONE. DURING TROUBLESHOOTING CUSTOMER HAD MOTOR ERROR ALARM DURING BOLUS DELIVERY. ADVISED CUSTOMER THAT INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
609384 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment