FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4130580 · Received September 30, 2014

Report

Report Number
3004209178-2014-95997
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
July 21, 2014
Report Date
July 21, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP ALARMED PRIME DURING BASIC OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED AND LOOSE DRIVE SUPPORT DISK. INSULIN PUMP HAD A CRACKED CASE ON DISPLAY WINDOW CORNER, MINOR SCRATCHES ON LCD WINDOW, CRACKED BATTERY TUBE THREADS AND CRACKED BELT CLIP SLOT ON BATTERY TUBE THREADS AREA.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER'S STATES INSULING SQUIRT DURING MANUAL PRIME FOLLOWED BY ALARM. BLANK SCREEN ALSO REPORTED. CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
609316 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWB

Patients

Seq Age Sex Outcome Treatment
1