PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-95989
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Date of Event
- July 21, 2014
- Report Date
- July 21, 2014
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
FAILURE ANALYSIS REVEALED THAT ALL THE BUTTONS ON THE INSULIN PUMP FUNCTIONING PROPERLY. NO UNLOCKED KEYPAD CONNECTOR NOTED DURING VISUAL INSPECTION. THE DEVICE WAS RECEIVED WITH SCRATCHED LCD WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
CUSTOMER'S FATHER CALLED STATING CUSTOMER IS NOT ALWAYS ABLE TO LOCK KEYPAD ON INSULIN PUMP. CUSTOMER WAS AWAY FOR THE HOLIDAYS, AND STATES WILL CALL BACK FOR REPLACEMENT INFORMATION.