FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4130553 · Received September 30, 2014

Report

Report Number
3004209178-2014-95971
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
July 21, 2014
Report Date
July 21, 2014
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS WERE NOTED. NO NUMBER RAMPING WAS NOTED ON THE DISPLAY. NO TRACES OF MOISTURE WERE NOTED AT THE ELECTRONIC OR MOTOR ASSEMBLIES PER VISUAL INSPECTION. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND MINOR SCRATCHES ON THE DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 341 MG/DL. THE CUSTOMER TREATED HER HIGH BLOOD GLUCOSE WITH A MANUAL INJECTION. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER ALSO DECLINED TO TROUBLESHOOT. CUSTOMER IS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
609604 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 51 YR