FDA Adverse Event Malfunction Summary report: N

EMAX 2 MOTOR

MDR report key: 4130430 · Received September 30, 2014

Report

Report Number
1045834-2014-12932
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
September 18, 2014
Report Date
September 18, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE HAD VIBRATION AND WAS RUNNING HOT. IT WAS DETERMINED THAT THE VIBRATION IS DUE TO THE BEARINGS BEING DESTROYED. IT WAS FURTHER DETERMINED THAT THE OVERHEATING WAS DUE TO A WORN MOTOR. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO THE DEVICE BEING WORN DUE TO NORMAL USE AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING ENGINEERING EVALUATION, IT WAS OBSERVED THAT THE MOTOR DEVICE WAS HAD VIBRATION AND WAS RUNNING HOT. IT WAS REPORTED THAT THIS EVENT IS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
609439 EMAX 2 MOTOR MOTOR, DRILL, ELECTRIC - HANDPIECE HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1