FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 4130424 · Received September 30, 2014

Report

Report Number
1416980-2014-33924
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
July 15, 2014
Report Date
September 5, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, AN INCREASED INTRA-PERITONEAL VOLUME EVENT WAS IDENTIFIED. THIS EVENT OCCURRED IN THE THERAPY THAT WAS INITIATED ON (B)(6) 2014 AT 09:20:47. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 749ML, INDICATING THE HOME PATIENT DRAINED 749ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1200ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
609437 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1