FDA Adverse Event Malfunction Summary report: N

INDIGO¿

MDR report key: 4130341 · Received September 30, 2014

Report

Report Number
1045254-2014-00236
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
September 5, 2014
Report Date
September 5, 2014
Manufacturer
MEDTRONIC XOMED INC.
Product Code
HBE
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CE
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IN RESPONSE TO MEDTRONIC¿S REQUEST FOR DEVICE RETURN, THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND REPAIR (DETAILED AS FOLLOWS): ANALYSIS WAS ABLE TO CONFIRM THE ALLEGED ISSUE, AS THE HEATRISE WAS 47 DEGREES F¿HOWEVER IT DID NOT RUN HOT DURING THE FIRST MINUTE OF TESTING BUT WITHIN 5 MINUTES. THE TEMPERATURE TESTS PRIOR TO REPAIR REVEALED THE FOLLOWING RESULTS IN THE TIMEFRAME OF ONE (1) MINUTE: T1: 96 DEGREES F ¿ T2: 79 DEGREES F; THEREFORE THE HOUSING WAS REPLACED AS A RESULT, ALONG WITH THE WORN MOTOR. METHOD: TEST FOR HIGH TEMPERATURE CONDITIONS.

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INDIGO HIGH SPEED OTOLOGIC DRILL ¿HEATED UP¿ INTRAOPERATIVELY IN A 3-HOUR OTOLOGY PROCEDURE ON A (B)(4) MALE, WEIGHING 150 POUNDS. THE HEAT UP OCCURRED IN UNDER A MINUTE, DESCRIBED AS ¿TRANSFERRING VERY FAST¿. THERE WAS NO REPORTED PATIENT/USER IMPACT OR INJURY AS A RESULT OF THIS EVENT. ANOTHER DRILL WAS USED TO SUCCESSFULLY COMPLETE THE SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
606999 INDIGO¿ DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MEDTRONIC XOMED INC. 1845000

Patients

Seq Age Sex Outcome Treatment
1 00033 YR