FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INFUSION PUMP

MDR report key: 4130335 · Received September 18, 2014

Report

Report Number
2032227-2014-24686
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
September 3, 2014
Report Date
September 3, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH A BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACES. THE DEVICE WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE, CRACKED RESERVOIR TUBE WINDOW AND MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTONS ERROR. THE BLOOD GLUCOSE READING WAS 129 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
578884 PARADIGM REAL-TIME INFUSION PUMP OYC MEDTRONIC MINIMED MMT-522LNAS

Patients

Seq Age Sex Outcome Treatment
1 46 YR