LIGASURE ADVANCE PISTOL GRIP
Report
- Report Number
- 1717344-2014-00674
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Date of Event
- July 3, 2014
- Report Date
- July 7, 2014
- Manufacturer
- COVIDIEN LP
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). TO DATE THE INCIDENT SAMPLE HAS NOT BEEN RECEIVED FOR EVAL. IF THE SAMPLE IS RECEIVED OR IF ADDITIONAL INFO PERTINENT TO THE INCIDENT IS OBTAINED A FOLLOW-UP REPORT WILL BE SUBMITTED.
(B)(4). DATE OF INITIAL REPORT : 08/01/2014. DATE OF FOLLOW-UP REPORT : 09/30/2014. THE CLEAR INSULATION WAS DAMAGED. THE DEVICE FAILED HIPOT TESTING. THE IFU STATES TO PREVENT DAMAGE TO THE FLEXIBLE INSULATION PROXIMAL TO THE JAWS, CONFIRM THE HANDLE IS FULLY CLOSED PRIOR TO INSERTION INTO AND EXTRACTION FROM THE CANNULA. USE THE APPROPRIATELY SIZED CANNULA TO ALLOW FOR EASY INSERTION AND EXTRACTION OF THE INSTRUMENT. FAILURE TO DO SO MAY IMPACT THE INTEGRITY OF THE FLEXIBLE INSULATION. CANNULAS WITH HARD, NON-BEVELED OPENINGS MAY CAUSE THE FLEXIBLE INSULATION TO RETRACT, WHICH MAY COMPROMISE THE INSULATION. IF RETRACTION OCCURS, THE INSTRUMENT MUST BE DISCARDED. DO NOT ATTEMPT TO CLEAN THE FLEXIBLE INSULATION. CLEANING MAY DAMAGE INSULATION.
THE CUSTOMER REPORTED THAT DURING A LAPAROSCOPIC NEPHRECTOMY, THE INSULATION AROUND THE JAWS WAS BUNCHING UP WHILE IN USE. THERE WAS NO PT INJURY. ANOTHER DEVICE OF THE SAME TYPE WAS OPENED AND USED TO CONTINUE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 608260 | LIGASURE ADVANCE PISTOL GRIP | LIGASURE VESSEL SEALING SYSTEM | GEI | COVIDIEN LP | 32760245X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |