FDA Adverse Event Malfunction Summary report: N

LIGASURE ADVANCE PISTOL GRIP

MDR report key: 4130319 · Received September 30, 2014

Report

Report Number
1717344-2014-00674
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
July 3, 2014
Report Date
July 7, 2014
Manufacturer
COVIDIEN LP
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). TO DATE THE INCIDENT SAMPLE HAS NOT BEEN RECEIVED FOR EVAL. IF THE SAMPLE IS RECEIVED OR IF ADDITIONAL INFO PERTINENT TO THE INCIDENT IS OBTAINED A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). DATE OF INITIAL REPORT : 08/01/2014. DATE OF FOLLOW-UP REPORT : 09/30/2014. THE CLEAR INSULATION WAS DAMAGED. THE DEVICE FAILED HIPOT TESTING. THE IFU STATES TO PREVENT DAMAGE TO THE FLEXIBLE INSULATION PROXIMAL TO THE JAWS, CONFIRM THE HANDLE IS FULLY CLOSED PRIOR TO INSERTION INTO AND EXTRACTION FROM THE CANNULA. USE THE APPROPRIATELY SIZED CANNULA TO ALLOW FOR EASY INSERTION AND EXTRACTION OF THE INSTRUMENT. FAILURE TO DO SO MAY IMPACT THE INTEGRITY OF THE FLEXIBLE INSULATION. CANNULAS WITH HARD, NON-BEVELED OPENINGS MAY CAUSE THE FLEXIBLE INSULATION TO RETRACT, WHICH MAY COMPROMISE THE INSULATION. IF RETRACTION OCCURS, THE INSTRUMENT MUST BE DISCARDED. DO NOT ATTEMPT TO CLEAN THE FLEXIBLE INSULATION. CLEANING MAY DAMAGE INSULATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING A LAPAROSCOPIC NEPHRECTOMY, THE INSULATION AROUND THE JAWS WAS BUNCHING UP WHILE IN USE. THERE WAS NO PT INJURY. ANOTHER DEVICE OF THE SAME TYPE WAS OPENED AND USED TO CONTINUE THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
608260 LIGASURE ADVANCE PISTOL GRIP LIGASURE VESSEL SEALING SYSTEM GEI COVIDIEN LP 32760245X

Patients

Seq Age Sex Outcome Treatment
1 UNK