FDA Adverse Event Malfunction Summary report: N

7.5CM ANGLE ATTACHMENT

MDR report key: 4130215 · Received September 30, 2014

Report

Report Number
1045834-2014-13114
Event Type
Malfunction
Date Received
September 30, 2014
Report Date
March 8, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. (B)(4). DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, IT WAS NOTICED THAT THE MECHANICS OF THE PUMP WERE NOT FUNCTIONING PROPERLY DUE TO NORMAL WEAR. THE DEVICE WAS REPLACED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE DEVICE GENERATED IMMENSE HEAT VERY QUICKLY AT THE ANGLED DRIVE COUPLING, AND BEARINGS IN THE ANGLED ATTACHMENT ARE WORN. THIS IS 1 OF 1 REPORT FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
607632 7.5CM ANGLE ATTACHMENT MOTOR, DRILL, ELECTRIC HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1