FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4130091 · Received September 30, 2014

Report

Report Number
1031452-2014-11631
Event Type
Malfunction
Date Received
September 30, 2014
Report Date
September 11, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

PER PROVIDER, ALARMING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
608224 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2

Patients

Seq Age Sex Outcome Treatment
1 Other