FDA Adverse Event Malfunction Summary report: N

OSCILL-SAW-ATTACHM W/QUICK-COUPL F/COMPA

MDR report key: 4130058 · Received September 30, 2014

Report

Report Number
2520274-2014-13742
Event Type
Malfunction
Date Received
September 30, 2014
Report Date
September 15, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
GFA
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SERIAL NUMBER WAS UNREADABLE. THEREFORE, THE MANUFACTURING LOCATION WAS UNKNOWN. THE DEVICE MANUFACTURE DATE IS UNKNOWN. (B)(6). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE:THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE WEAR FROM NORMAL USE AND SERVICING OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED FORM GERMANY THAT THE WHEN IN USE, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE, WAS NOISY AND DID NOT KEEP THE SAW BLADE LOCKED INTO PLACE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. THERE WERE NO DELAYS TO A SCHEDULED SURGICAL PROCEDURE. A SPARE DEVICE WAS NOT AVAILABLE FOR USE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
607690 OSCILL-SAW-ATTACHM W/QUICK-COUPL F/COMPA BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL GFA DEPUY SYNTHES POWER TOOLS NI

Patients

Seq Age Sex Outcome Treatment
1