STAPLE, IMPLANTABLE
Report
- Report Number
- 3005075853-2014-06762
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Date of Event
- August 11, 2014
- Report Date
- August 19, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. RELEASE BUTTON THE ANALYSIS FOUND THAT ONE PLE60A DEVICE WAS RECEIVED FOR ANALYSIS WITHOUT CARTRIDGE RELOAD. THE MANUAL OVERRIDE DOOR WAS NOTED TO BE OUT OF POSITION AND MISSING; THE OVERRIDE LEVER WAS UP WHICH DENOTES THAT THE KNIFE WAS MANUALLY RETURNED TO HOME POSITION. PLEASE NOTE IF THE INSTRUMENT IS PARTIALLY FIRED, SLIDE THE KNIFE REVERSE SWITCH FORWARD TO RETURN THE KNIFE TO HOME POSITION. IT SHOULD BE NOTED AFTER THE MANUAL OVERRIDE SYSTEM IS USED THE INSTRUMENT IS DISABLED AND CANNOT BE USED FOR ANY SUBSEQUENCE FIRINGS. FURTHERMORE, THE DEVICE WAS FOUND TO HAVE THE CLAMPING MECHANISM DAMAGED. DUE TO THE WEAR AND DAMAGE TO THE CLAMP RELEASE, IT APPEARS THAT THE DEVICE WAS FORCED OPEN WITH THE KNIFE NOT AT THE HOME POSITION. THE DEVICE WAS DISASSEMBLED TO RESET THE BAILOUT SYSTEM AND THEN, THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE DEVICE CLOSED, CYCLED AND OPENED WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE DID NOT FIRE AT THE FIRST FIRING. IT IS UNKNOWN WHAT COLOR CARTRIDGE WAS BEING USED. A SECOND DEVICE WAS PULLED AND THE SAME THING OCCURRED. A THIRD DEVICE WAS PULLED TO COMPLETE THE PROCEDURE WITH NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 608800 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | L4EY7Y |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ASKU |