FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4130038 · Received September 19, 2014

Report

Report Number
2032227-2014-26020
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
July 23, 2014
Report Date
July 23, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, BROKEN BATTERY TUBE THREADS, CRACKED CASE ON THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND STAINED END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THE EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN CASING WAS DAMAGED AND THAT THE RESERVOIR COULD NOT BE PROPERLY LOADED. THE INSULIN PUMP HAD ALSO ALARMED FOR NO DELIVERY. NO BLOOD GLUCOSE READING WAS GIVEN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
585173 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC MEDTRONIC MINIMED MMT-754WWP

Patients

Seq Age Sex Outcome Treatment
1