FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4123752 · Received September 26, 2014

Report

Report Number
3004209178-2014-90447
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
July 6, 2014
Report Date
July 6, 2014
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO UNEXPECTED NUMBER RAMPING WAS NOTED DURING TESTING. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE BUTTONS ON THE KEYPAD DOES NOT WORK. CUSTOMER'S BLOOD GLUCOSE AT THE TIME WAS 190MG/DL. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL BE REPLACED. NO FURTHER INFORMATION PROVIDED.