PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-91305
- Date Received
- September 26, 2014
- Date of Event
- July 9, 2014
- Report Date
- July 9, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. NO DISPLAY RAMPING ON ITS OWN ANOMALY NOTED. INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS AND MINOR SCRATCHES ON DISPLAY WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT IS REPORTED THAT THE PATIENT HAS A BUTTON ERROR ON THEIR INSULIN PUMP AND THE BUTTON WILL NOT WORK. PATIENT'S BLOOD GLUCOSE LEVEL IS AT 95MG/DL. TROUBLE SHOOTING IS CONDUCTED AND FOUND WATER DAMAGE. DEVICE IS BEING RETURNED. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 602910 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | MEDTRONIC MINIMED | MMT-723LNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |