FDA Adverse Event Summary report: N

PARADIGM REAL-TIME REVEL INSUILN INFUSION PUMP

MDR report key: 4123707 · Received September 26, 2014

Report

Report Number
3004209178-2014-89259
Date Received
September 26, 2014
Date of Event
July 3, 2014
Report Date
July 3, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATION CO
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP WAS RECEIVED WITH BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED DISPLAY WINDOW CORNER, CRACKED BELT CLIP SLOT AT BATTERY TUBE THREADS AREA AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED BUTTON ERROR ALARM ON HIS INSULIN PUMP. TROUBLESHOOTING FOUND ON EVENTS LEADING TO THE ALARM. CUSTOMER IS ADVISED TO REVERT TO HIS BACK-UP PLAN WHILE HIS PUMP IS BEING REPLACED. CUSTOMER STATED HE DID NOT HAVE A BACK-UP PLAN AND IS ADVISED TO CONTACT HIS PREVIOUS DOCTOR TO CREATE A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602378 PARADIGM REAL-TIME REVEL INSUILN INFUSION PUMP OYC MEDTRONIC PUERTO RICO OPERATION CO MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 24 YR