G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2014-25536
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- August 30, 2014
- Report Date
- August 30, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. AN ALARM TEXT AND INTERIOR INSPECTION WAS PERFORMED AND THE DEVICE FAILED, CONFIRMING THE REPORTED EVENT OF LOW AUDIO OUTPUT. THE ROOT CAUSE WAS DETERMINED TO BE AN ASSEMBLY ERROR.
(B)(4).
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM LOW AUDIO OUTPUT ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, AT THE TIME OF THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT TO TEST THE ALERT FUNCTIONALITY AND PATIENT REPORTED ALERTS WERE FUNCTIONAL.