FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4123669 · Received September 26, 2014

Report

Report Number
2032227-2014-29813
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 29, 2014
Report Date
August 29, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING NOTED. THE INSULIN PUMP HAS CRACKED CASE AT DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS, RESERVOIR TUBE LIP, BROKEN BELT CLIP SLOT, MINOR SCRATCHED AND CRACKED DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ON HER INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE WAS 342 MG/DL. SHE WAS TRYING TO BOLUS FOR HER HIGH BLOOD GLUCOSE WHEN THE DEVICE ALARMED BUTTON ERROR. CUSTOMER WAS ADVISED THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
599977 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 58 YR