FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4123641 · Received September 26, 2014

Report

Report Number
2032227-2014-29797
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 28, 2014
Report Date
August 29, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED FOR A BUTTON ERROR AND HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE OCCLUSION TEST COULD NOT BE PERFORMED DUE TO A KEYPAD ANOMALY. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT A DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP, AND MINOR SCRATCHES ON THE DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD AN UNRESPONSIVE BATTERY, AND IT ALARMED BUTTON ERROR AFTER THE CUSTOMER REPLACED THE BATTERY IN AN ATTEMPT TO RESOLVE THE ISSUE. THE BLOOD GLUCOSE READING WAS 200 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600192 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAH

Patients

Seq Age Sex Outcome Treatment
1 35 YR