PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-29620
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- August 18, 2014
- Report Date
- August 28, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).
IT WAS REPORTED BY CUSTOMER THAT THEY HAD BEEN HOSPITALIZED FOR LOW BLOOD GLUCOSE LEVELS ON (B)(6) OF 2014. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 20 MG/DL AT TIME OF HOSPITALIZATION AND WAS GIVEN INTRAVENOUS INSULIN DRIP WHILE IN THE HOSPITAL. DURING TROUBLESHOOTING, CUSTOMER WAS ASKED TO CHECK THE CONDITION OF THE DRIVE SUPPORT CAP ON THE DEVICE AND FOUND IT TO BE NORMAL. NEXT, CUSTOMER WAS ASKED TO CHECK FOR AIR BUBBLES IN INFUSION SET TUBING. CUSTOMER VERIFIED THERE WERE NO AIR BUBBLES. AFTERWARDS, THE RESERVOIR WAS REINSERTED INTO INSULIN PUMP AND A MANUAL PRIME WAS PERFORMED. DEVICE PASSED PRIMING, HIGH PRESSURE TEST AND INSULIN EXITED TUBING. INFORMED CUSTOMER DEVICE IS FUNCTIONING PROPERLY. ADVISED TO CHANGE ENTIRE SET AND MONITOR DEVICE. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 600401 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Hospitalization |