FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4123442 · Received September 26, 2014

Report

Report Number
2531779-2014-27695
Event Type
Malfunction
Date Received
September 26, 2014
Report Date
September 16, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/29/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP BLACK BOX DATA REVEALED EVIDENCE OF PARTIALLY DISCHARGED BATTERY INSTALLATIONS. DURING A VISUAL INSPECTION OF THE PUMP, NO DAMAGE WAS FOUND TO THE BATTERY COMPARTMENT OR RETURNED BATTERY CAP. THE PUMP POWERED ON PROPERLY WITH THE RETURNED BATTERY CAP. CURRENT DRAWS MEASURED WITHIN SPECIFICATIONS. THE PUMP CASE WAS REMOVED, AND NO EVIDENCE OF MOISTURE INGRESS OR LOOSE COMPONENTS WAS FOUND.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600087 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 29 YR