FDA Adverse Event Injury Summary report: N

SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

MDR report key: 4123426 · Received September 26, 2014

Report

Report Number
1416980-2014-33498
Event Type
Injury
Date Received
September 26, 2014
Date of Event
September 1, 2014
Report Date
September 2, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
KDJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT NO: CMPLNT-(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). AS THE SAMPLE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, IT WAS NOTED THAT PATIENT LINE EXTENSIONS WERE CONNECTED AFTER PRIMING. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. IT ALSO INSTRUCTS THE USER TO CONNECT THE PATIENT LINE EXTENSION TO THE PATIENT LINE PRIOR TO PRIMING. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT CONNECTED THEIR PATIENT EXTENSION LINE AFTER PRIMING ON A HOMECHOICE DEVICE AND PERFORMED THERAPY, RESULTING IN THE PATIENT GETTING AIR IN THEIR SHOULDERS. THE PATIENT WAS HOSPITALIZED FOR THE EVENT. TREATMENT FOR THE EVENT WAS NOT REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600316 SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 45 YR Hospitalization HOMECHOICE