FDA Adverse Event Injury Summary report: N

TALENT

MDR report key: 4123417 · Received September 26, 2014

Report

Report Number
2953200-2014-01949
Event Type
Injury
Date Received
September 26, 2014
Date of Event
August 15, 2014
Report Date
September 3, 2014
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P070027
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

A TALENT AAA STENT GRAFT SYSTEM WAS IMPLANTED FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. ANEURYSM AND VESSEL MORPHOLOGY AT IMPLANT ARE UNKNOWN. IT WAS REPORTED THAT THE PATIENT WAS SUFFERING FROM BACK PAIN AND SOUGHT RELIEF. ON A FOLLOW UP CT SCAN, THE BIFURCATE HAD MIGRATED INTO THE ANEURYSM SAC CAUSING A PROXIMAL TYPE I ENDOLEAK. THE BIFURCATED DEVICE WAS NOT SALVAGEABLE VIA ENDOVASCULAR MEANS AS THE PROXIMAL SUPRARENAL CROWNS HAD INCORPORATED INTO THE THROMBUS PRESENT IN THE DISEASED AORTA. THE GRAFT HAD LAID OVER IN THE AORTA APPROXIMATELY 85 DEGREES FROM ITS ORIGINAL IMPLANTED POSITION THUS CONTRIBUTING TO THE PHYSICIANS DECISION TO EXPLANT THE GRAFT. A DACRON GRAFT WAS SEWN INTO THE PATIENT TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600313 TALENT SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC IRELAND V00385964

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention