FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 4123326 · Received September 26, 2014

Report

Report Number
2024168-2014-06256
Event Type
Injury
Date Received
September 26, 2014
Date of Event
September 3, 2014
Report Date
September 3, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED SUTURE DETACHMENT COULD NOT BE CONFIRMED. THE ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE SUTURE WAS RETURNED OUTSIDE THE DEVICE AND THE PRE-TIED KNOT HAD BEEN FORMED, FULLY ADVANCED AND TIGHTENED. BASED ON ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES FOR THIS LOT. A QUERY OF THE ELECTRONIC COMPLAINT HANDLING DATABASE REVEALED NO OTHER INCIDENTS FOR SUTURE DETACHMENT REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO EVIDENCE TO INDICATE THE PRESENCE OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). A FEMORAL VENOUS SHEATH WAS REPORTED TO BE NEXT TO THE ARTERIAL SHEATH DURING CLOSING. PER THE INSTRUCTIONS FOR USE, THE SAFETY AND EFFECTIVENESS OF THE PROGLIDE DEVICE HAVE NOT BEEN ESTABLISHED FOR PATIENTS WITH IPSILATERAL FEMORAL VENOUS SHEATH DURING THE CATHETERIZATION PROCEDURE. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 5F SHEATH AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, WHEN ADVANCING THE KNOT WITH THE KNOT PUSHER THE SUTURE BROKE. THE PHYSICIAN REPORTED NO FORCE WAS USED. HEMOSTASIS WAS ACHIEVED USING MANUAL ARTERIAL COMPRESSION. A FEMORAL VENOUS SHEATH WAS REPORTED TO BE NEXT TO ARTERIAL SHEATH DURING CLOSING. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. THE PHYSICIAN IS REPORTED TO BE IN-TRAINING IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602510 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 30903K1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 5F