PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4123220 · Received September 26, 2014

Report

Report Number
2032227-2014-29478
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP PASSED BASIC OCCLUSION TEST AND DISPLACEMENT TEST. NO MOTOR ERROR ALARMS WERE NOTED. HOWEVER, INSULIN PUMP WAS UNABLE TO PRIME UNIT DURING PRIME TEST DUE TO FAULTY FORCE SENSOR. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, BATTERY TUBE THREADS AND MINOR SCRATCHES ON DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURIN BOLUS DELIVERY. CUSTOMER'S BLOOD GLUCOSE WAS 327 MG/DL. TROUBLESHOOTING WAS DONE. ADVISED CUSTOMER THE INSULIN PUMP WOULD BE REPLACED. ADVISED TO DISCONTINUE USING THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROVIDER. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602467 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAB

Patients

Seq Age Sex Outcome Treatment
1 9 YR