FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4123195 · Received September 26, 2014

Report

Report Number
2032227-2014-28864
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE KEYPAD TRACE. NO BUTTON ERROR ALARM. THE INSULIN PUMP HAS MINOR SCRATCHED LCD WINDOW, CRACKED CASE NEAR DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS CRACKED AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE READING WAS 180 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602433 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LNAB

Patients

Seq Age Sex Outcome Treatment
1 28 YR