FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4123179 · Received September 26, 2014

Report

Report Number
2032227-2014-28838
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO UNEXPECTED BUTTON ERROR ALARMS NOTED. INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES NOTED. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE AT LCD WINDOW CORNERS, BROKEN RESERVOIR TUBE LIP, SCRATCHES ON RESERVOIR TUBE WINDOW, CRACKED BATTERY TUBE THREADS, BROKEN BELT CLIP SLOT, MISSING END CAP STICKER AND STAINED ADDRESS AND SERIAL NUMBER LABEL. SLANTED AND LOOSE DRIVE SUPPORT DISK NOTED. INSULIN PUMP PASSED DISPLACEMENT TEST, REWIND TEST, BASIC OCCLUSION, OCCLUSION TEST AND EXCESSIVE NO DELIVERY TEST. INSULIN PUMP FAILED PRIME TEST DUE TO BAD FORCE SENSOR.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602918 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAP

Patients

Seq Age Sex Outcome Treatment
1 67 YR