FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4123162
·
Received September 26, 2014
Report
- Report Number
- 2032227-2014-28848
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- August 28, 2014
- Report Date
- August 28, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS WERE NOTED DURING THE VISUAL INSPECTION. NO BUTTON ERROR ALARMS WERE NOTED AND ALL OF THE BUTTONS FUNCTIONED PROPERLY. HOWEVER, MOISTURE DAMAGE WAS NOTED ON THE KEYPAD TRACES.
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 220MG/DL. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 602638 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |