FDA Adverse Event Malfunction Summary report: N

XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 4123133 · Received September 26, 2014

Report

Report Number
2024168-2014-06251
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
September 2, 2014
Report Date
September 2, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE FAILURE TO ADVANCE AND DIFFICULTY REMOVING THE DEVICE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON A VISUAL AND DIMENSIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY/REVIEW OF THE COMPLAINT DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A HEAVILY CALCIFIED, 90% STENOSED, PROXIMAL RIGHT CORONARY ARTERY PROCEDURE, THE 3.5 X 15 MM XIENCE PRIME STENT FAILED TO CROSS THE LESION DUE TO THE PATIENTS ANATOMY. THERE WAS RESISTANCE DURING ADVANCEMENT AND REMOVAL WITH THE VESSEL. ANOTHER XIENCE PRIME WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS OR A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE REPORTED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601153 XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 405094N

Patients

Seq Age Sex Outcome Treatment
1 71 YR