FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4123114 · Received September 26, 2014

Report

Report Number
2032227-2014-29465
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD A BLANK DISPLAY DUE TO A MISSING BATTERY TUBE SPRING. THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, AND EXCESSIVE NO DELIVERY ALARM TESTS COULD NOT BE PERFORMED DUE TO THE BLANK DISPLAY. THE INSULIN PUMP WAS RECEIVED WITH CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP. THE BATTERY TUBE WAS CORRODED.

Description of Event or Problem · 1

IT IS REPORTED THAT A CUSTOMER RECEIVED A BLANK DISPLAY SCREEN ON THEIR INSULIN PUMP. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS AT 290 MG/DL. THE CUSTOMER STATED THAT THERE WERE NO SIGNIFICANT EVENTS THAT COULD HAVE LED TO THE ISSUE. THE CUSTOMER WAS INFORMED THAT THE PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. A NEW PUMP WAS SHIPPED TO THE CUSTOMER. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600476 530G INSULIN PUMP OZO OZO MEDTRONIC MINIMED MMT-551NAL

Patients

Seq Age Sex Outcome Treatment
1 44 YR