FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4123110 · Received September 26, 2014

Report

Report Number
2032227-2014-29460
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL. THE DISPLACEMENT TEST COULD NOT BE PERFORMED DUE TO THESE ALARMS. A CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS WERE NOTED DURING THE VISUAL INSPECTION.

Description of Event or Problem · 1

IT IS REPORTED THAT A CUSTOMER RECEIVED AN ALARM ON THEIR INSULIN PUMP. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS AT 400 MG/DL. THE CUSTOMER STATED THAT THE PUMP WAS IN THE SAME ROOM AS A MRI WHICH COULD HAVE LED TO THE ISSUE. THE CUSTOMER WAS INFORMED THAT THE PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. A NEW PUMP WAS SHIPPED TO THE CUSTOMER. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600755 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 81 YR