PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-29533
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- August 11, 2014
- Report Date
- August 28, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).
INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST, DUE TO PROTRUDED AND LOOSE DRIVE SUPPORT DISK. NO MOTOR ERROR ALARM. MOTOR PASSED MOTOR TEST. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE NEAR DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, BROKEN BELT CLIP SLOT AND CRACKED RESERVOIR TUBE LIP.
CUSTOMER REPORTED A MOTOR ERROR DURING BOLUS ON HER INSULIN PUMP FOR THE LAST THREE WEEKS. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. THERE WERE NO SIGNIFICANT EVENTS LEADING TO THE ALARM. THE DEVICE WAS NOT EXPOSED TO A STRONG MAGNETIC FIELD. THE ALARM WAS CLEARED. CUSTOMER DOES NOT USE THE SENSOR FEATURE. CUSTOMER IS ABLE TO REWIND THE DEVICE. SHE WAS ADVISED THAT THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 600734 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |