FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 4123000 · Received September 26, 2014

Report

Report Number
2032227-2014-29543
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP WAS ALARMING NO DELIVERY. CUSTOMER STATES THAT THEY HAVE CHANGED THE RESERVOIR AND INFUSION SET, YET ISSUE HAS CONTINUED. DURING TROUBLESHOOTING, CUSTOMER WAS ASKED TO PERFORM A FIXED PRIME OF INSULIN. CUSTOMER STATES THE INSULIN DID EXIT. NEXT, CUSTOMER WAS ASKED TO MANUAL PRIME OF INSULIN. CUSTOMER STATES THE INSULIN DID EXIT AND EXPERIENCED GREATER THAN NORMAL RESISTANCE. ADVISED CUSTOMER TO CHANGE RESERVOIR AND SET. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 103 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600723 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC MINIMED MMT-326A

Patients

Seq Age Sex Outcome Treatment
1 53 YR