LIGAMAX 5MM
Report
- Report Number
- 3005075853-2014-06696
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- August 26, 2014
- Report Date
- August 28, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOR THE EL5ML DEVICE FOUND THAT IT WAS RETURNED WITH THE SHAFT BENT. A BAG WITH ONE MALFORMED CLIP WAS RECEIVED ALONG WITH THE INSTRUMENT. NO FUNCTIONAL TESTING COULD BE PERFORMED DUE TO THE RETURNED CONDITION OF THE DEVICE. POSSIBLE CAUSES FOR THE DAMAGE FOUND MAY BE INADVERTENT PRESSURE PLACED ON THE DEVICE SHAFT THROUGH BENDING, PRESSING AGAINST THE TROCAR, OTHER DEVICES ON THE BACK TABLE, USING THE DEVICE AS A RETRACTOR OR MOVING HEAVY TISSUE WITH THE DEVICE SHAFT. HOWEVER, IT COULD NOT BE DETERMINED WHAT TO MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE JAMMED WITH THE JAWS CLOSED WHEN DEPLOYING A CLIP. THE JAWS EVENTUALLY OPENED AND ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITHOUT INCIDENT. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601210 | LIGAMAX 5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |