FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4122949 · Received September 26, 2014

Report

Report Number
2032227-2014-28810
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD AN UNRESPONSIVE ACT BUTTON DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER REPORTED UNRESPONSIVE BUTTONS ON HER INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS 340 MG/DL. CUSTOMER TREATED HER HIGH BLOOD GLUCOSE WITH A BOLUS DELIVERY. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE KEYPAD ISSUES. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601191 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 40 YR