PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-28778
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- August 28, 2014
- Report Date
- August 28, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. (B)(4).
INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP HAD A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.
IT WAS REPORTED THAT THE BUTTONS WERE INTERMITTENTLY OR NOT RESPONDING ON THE CUSTOMER'S INSULIN PUMP. THE INSULIN PUMP REPORTEDLY WAS NOT DROPPED, OR EXPOSED TO MOISTURE. THE CUSTOMER'S BLOOD GLUCOSE WAS 7.2 MMOL/L.. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 600803 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-754WWS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |