FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4122904 · Received September 26, 2014

Report

Report Number
2032227-2014-28767
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
May 1, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT NO BUTTON ERROR ALARM NOTED. HOWEVER, THE ESCAPE BUTTON DID NOT RESPOND DUE TO FLATTEN BUTTON DOME SWITCH. UNABLE TO PERFORM THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST DUE TO THE UNRESPONSIVE BUTTON. THE DEVICE WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTS TO HAVE RECEIVED A BUTTON ERROR ALARM. CUSTOMER STATES THAT THE BUTTONS WERE PUSHED IN CAUSING SCREEN TO MOVE WITHOUT PRESSING ON BUTTONS. CUSTOMER'S BLOOD GLUCOSE AT THE TIME OF CALL WAS 326 MG/DL WHICH WAS TREATED WITH MANUAL INJECTION. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. NO FURTHER INFORMATION PROVIDED.