FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4122762 · Received September 18, 2014

Report

Report Number
2032227-2014-24822
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
August 19, 2014
Report Date
August 19, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP HAD A CRACKED CASE NEAR THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE DISPLAY WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND A CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT IS REPORTED THAT A CUSTOMER RECEIVED A MOTOR ERROR ALARM ON THEIR INSULIN PUMP WHILE SHE WAS TRYING TO BOLUS. THE PT'S BLOOD GLUCOSE LEVEL WAS AT 130 MG/DL. THE CUSTOMER STATED THAT THERE WERE NO SIGNIFICANT EVENTS THAT COULD HAVE LED TO THE MOTOR ERROR ALARM. THE CUSTOMER WAS ADVISED THAT THEY CAN REDUCE THE LIKELIHOOD OF THESE OCCURRENCES BY ALLOWING THE BOLUS DELIVERY TO COMPLETE BEFORE ACCESSING THE SENSOR GLUCOSE GRAPH. THE CUSTOMER WAS INFORMED THAT THE PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. A NEW PUMP WAS SHIPPED TO THE CUSTOMER. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
579431 530G INSULIN PUMP OZO MEDTRONIC MINIMED MMT-751NAP

Patients

Seq Age Sex Outcome Treatment
1 47 YR