530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-24822
- Event Type
- Malfunction
- Date Received
- September 18, 2014
- Date of Event
- August 19, 2014
- Report Date
- August 19, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP HAD A CRACKED CASE NEAR THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE DISPLAY WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND A CRACKED RESERVOIR TUBE LIP.
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
IT IS REPORTED THAT A CUSTOMER RECEIVED A MOTOR ERROR ALARM ON THEIR INSULIN PUMP WHILE SHE WAS TRYING TO BOLUS. THE PT'S BLOOD GLUCOSE LEVEL WAS AT 130 MG/DL. THE CUSTOMER STATED THAT THERE WERE NO SIGNIFICANT EVENTS THAT COULD HAVE LED TO THE MOTOR ERROR ALARM. THE CUSTOMER WAS ADVISED THAT THEY CAN REDUCE THE LIKELIHOOD OF THESE OCCURRENCES BY ALLOWING THE BOLUS DELIVERY TO COMPLETE BEFORE ACCESSING THE SENSOR GLUCOSE GRAPH. THE CUSTOMER WAS INFORMED THAT THE PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. A NEW PUMP WAS SHIPPED TO THE CUSTOMER. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 579431 | 530G INSULIN PUMP | OZO | MEDTRONIC MINIMED | MMT-751NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |