FDA Adverse Event Malfunction Summary report: N

QD11-G1_AO

MDR report key: 4122688 · Received September 26, 2014

Report

Report Number
1045834-2014-12891
Event Type
Malfunction
Date Received
September 26, 2014
Report Date
September 8, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
ERL
PMA / PMN Number
PK133604
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE (B)(4) EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE MET ALL MANUFACTURE SPECIFICATIONS. THEREFORE, THE REPORTED CONDITION COULD NOT BE CONFIRMED. AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.

Description of Event or Problem · 1

THIS IS REPORT 1 OF 2 FOR THE SAME EVENT. IT WAS REPORTED THAT WHILE IN USE WITH AN ATTACHMENT DEVICE, THE MOTOR DEVICE GOT SO HOT THAT IT ¿BURNED THROUGH THE DRAPE WHEN IT WAS SET DOWN¿. IT WAS FURTHER REPORTED THAT THE DEVICES WERE IN USE WITH A ¿MATCHSTICK¿ CUTTER DEVICE. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601464 QD11-G1_AO DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1 CUTTER DEVICE| MOTOR DEVICE