FDA Adverse Event Injury Summary report: N

CONTOUR NEXT

MDR report key: 4122595 · Received September 26, 2014

Report

Report Number
1826988-2014-00326
Event Type
Injury
Date Received
September 26, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K110894
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IN SOME COUNTRIES OUTSIDE THE US, CUSTOMER INFORMATION, IS NOT PROVIDED DUE TO PRIVACY LAWS.

Description of Event or Problem · 1

AN ADVOCATE FROM (B)(6) STATED HIS WIFE HAD HYPOGLYCEMIC SYMPTOMS BUT THE CONTOUR NEXT LINK GAVE HER A BLOOD GLUCOSE READING OF 127MG/DL. HE CALLED AN AMBULANCE AND THE PATIENT WAS RETESTED ON A DIFFERENT METER WITH A READING OF 20MG/DL. THE PARAMEDIC GAVE HER FOUR GLUCOSE UNITS. THE PATIENT WAS TAKEN TO THE HOSPITAL AND A BLOOD TEST WAS PERFORMED ON HER. THE READING WAS 64MG/DL. INFORMATION REGARDING FURTHER TREATMENT WAS NOT PROVIDED. SHE WAS ULTIMATELY DISCHARGED AND SENT HOME. NO ADDITIONAL INFORMATION REGARDING THE EVENT WAS PROVIDED. THE TEST STRIPS ARE EXPECTED TO BE RETURNED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601009 CONTOUR NEXT BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE LLC 4AFED03

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R