FDA Adverse Event
Injury
Summary report: N
CONTOUR NEXT
MDR report key: 4122592
·
Received September 26, 2014
Report
- Report Number
- 1826988-2014-00325
- Event Type
- Injury
- Date Received
- September 26, 2014
- Date of Event
- September 4, 2014
- Report Date
- September 16, 2014
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K121190
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
.
Description of Event or Problem · 1
A CANADIAN CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF 26.8MMOL/L ON THE CONTOUR NEXT METER. SHE FELT DIZZY AND NAUSEATED SO CALLED AN AMBULANCE. THE EMERGENCY PERSONNEL RETESTED HER ON A DIFFERENT METER AND THE READING WAS 8.6MMOL/L. THE CUSTOMER WAS TAKEN TO THE HOSPITAL, BUT COULD NOT PROVIDE ANY ADDITIONAL INFORMATION REGARDING THE INCIDENT. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION. NEW TEST STRIPS AND METER WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601008 | CONTOUR NEXT | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 3JFEC32 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |