PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-24547
- Event Type
- Malfunction
- Date Received
- September 18, 2014
- Date of Event
- August 16, 2014
- Report Date
- August 19, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TOT HE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. CURRENTLY IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACES. THE DEVICE HAD MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNER, BROKEN BELT CLIP SLOT AT BATTERY TUBE THREADS AREA AND CRACKED RESERVOIR TUBE LIP.
IT WAS REPORTED THAT TH INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 160 MG/DL. NO FURTHER INFO PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 579243 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | MEDTRONIC MINIMED | MMT-754CAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |