FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4122511 · Received September 22, 2014

Report

Report Number
1052693-2014-00360
Event Type
Malfunction
Date Received
September 22, 2014
Date of Event
August 29, 2014
Report Date
December 17, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT NOT YET RETURNED.

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF ABOUT HIGH BLOOD RESULTS. CUSTOMER NORMAL RESULTS ARE IN THE 160MG/DL RANGE. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE HIGHEST RESULT IN MEMORY (424) AND THE LOWEST NORMAL RESULT (160) IS LOCATED IN ZONE C. NO ADVERSE EVENT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF ABOUT HIGH BLOOD RESULTS. CUSTOMER NORMAL RESULTS ARE IN THE 160MG/DL RANGE. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE HIGHEST RESULT IN MEMORY (424) AND THE LOWEST NORMAL RESULT (160) IS LOCATED IN ZONE C. NO ADVERSE EVENT REPORTED.