FDA Adverse Event
Malfunction
Summary report: N
TRUERESULT
MDR report key: 4122511
·
Received September 22, 2014
Report
- Report Number
- 1052693-2014-00360
- Event Type
- Malfunction
- Date Received
- September 22, 2014
- Date of Event
- August 29, 2014
- Report Date
- December 17, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PRODUCT NOT YET RETURNED.
Additional Manufacturer Narrative · 1
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.
Description of Event or Problem · 1
CONSUMER COMPLAINT OF ABOUT HIGH BLOOD RESULTS. CUSTOMER NORMAL RESULTS ARE IN THE 160MG/DL RANGE. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE HIGHEST RESULT IN MEMORY (424) AND THE LOWEST NORMAL RESULT (160) IS LOCATED IN ZONE C. NO ADVERSE EVENT REPORTED.
Description of Event or Problem · 1
CONSUMER COMPLAINT OF ABOUT HIGH BLOOD RESULTS. CUSTOMER NORMAL RESULTS ARE IN THE 160MG/DL RANGE. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE HIGHEST RESULT IN MEMORY (424) AND THE LOWEST NORMAL RESULT (160) IS LOCATED IN ZONE C. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 587793 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PP1745 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |