TRUERESULT
Report
- Report Number
- 1052693-2014-00344
- Event Type
- Malfunction
- Date Received
- September 22, 2014
- Date of Event
- August 5, 2014
- Report Date
- December 17, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT NOT YET RETURNED FOR EVAL. (B)(4). MFR ACKNOWLEDGES LATENESS OF THE REPORT, PER INTERNAL CORRECTIVE ACTION. THIS REPORT SHOULD HAVE BEEN IDENTIFIED AS REPORTABLE WITHIN 30 DAYS OF THE IDENTIFICATION ((B)(4) 2014).
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S MISUNDERSTANDING OF ERRATIC RESULTS.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CALLER NORMALLY IS IN THE 100MG/DL'S. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE LOWEST RESULT GIVEN (60) AND THE NORMAL RESULT (100) IS LOCATED IN ZONE B/C. NO ADVERSE EVENT REPORTED.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CALLER NORMALLY IS IN THE 100 MG/DL'S. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE LOWEST RESULT GIVEN (60) AND THE NORMAL RESULT (100) IS LOCATED IN ZONE B/C. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 587621 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PP1678 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |