FDA Adverse Event
Malfunction
Summary report: N
TERUMO TCM II COOLING AND HEATING SYSTEM
MDR report key: 4122505
·
Received September 22, 2014
Report
- Report Number
- 1828100-2014-00765
- Event Type
- Malfunction
- Date Received
- September 22, 2014
- Date of Event
- August 31, 2014
- Report Date
- August 31, 2014
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- K883603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
PER TECHNICAL SUPPORT TO THE FIELD SERVICE REP (FSR): "IT SOUNDS LIKE THE MEMBRANE PANEL OR ONE OF THE DISPLAY BOARDS MAY BE SHORTED. YOU CAN OHM THE MEMBRANE PANEL FOR SHORTS AFTER DISCONNECTING IT FROM THE BOARDS. IF IT IS NOT EITHER THE MEMBRANE PANELS OR DISPLAY BOARDS, IT COULD BE THE COMPUTER CARD."
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE USE OF THE DEVICE FOR A NON-CLINICAL ACTIVITY, THE PERFUSIONIST (CCP) WAS SANITIZING THE COOLER HEATER UNIT LAST EVENING. THE UNIT WORKED AT FIRST AND TEN MINUTES LATER, HE WAS UNABLE TO MAKE ANY MODE SELECTIONS ON THE CONTROL PANEL. THE CCP SAID EVEN THE CARDIOPLEGIA SELECTIONS WOULD NOT WORK. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 587651 | TERUMO TCM II COOLING AND HEATING SYSTEM | TERUMO TCM II COOLING AND HEATING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 4415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |