FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 4122505 · Received September 22, 2014

Report

Report Number
1828100-2014-00765
Event Type
Malfunction
Date Received
September 22, 2014
Date of Event
August 31, 2014
Report Date
August 31, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PER TECHNICAL SUPPORT TO THE FIELD SERVICE REP (FSR): "IT SOUNDS LIKE THE MEMBRANE PANEL OR ONE OF THE DISPLAY BOARDS MAY BE SHORTED. YOU CAN OHM THE MEMBRANE PANEL FOR SHORTS AFTER DISCONNECTING IT FROM THE BOARDS. IF IT IS NOT EITHER THE MEMBRANE PANELS OR DISPLAY BOARDS, IT COULD BE THE COMPUTER CARD."

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE USE OF THE DEVICE FOR A NON-CLINICAL ACTIVITY, THE PERFUSIONIST (CCP) WAS SANITIZING THE COOLER HEATER UNIT LAST EVENING. THE UNIT WORKED AT FIRST AND TEN MINUTES LATER, HE WAS UNABLE TO MAKE ANY MODE SELECTIONS ON THE CONTROL PANEL. THE CCP SAID EVEN THE CARDIOPLEGIA SELECTIONS WOULD NOT WORK. THERE WAS NO PT INVOLVEMENT.