PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4122487 · Received September 18, 2014

Report

Report Number
2032227-2014-24554
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
August 19, 2014
Report Date
August 19, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD NO UP ARROW BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, A CRACKED CASE AT DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE DISPLAY WINDOW, AND SCRATCHES ON THE RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD A BUTTON ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 153 MG/DL. NO FURTHER INFO REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
579242 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC MEDTRONIC MINIMED MMT-523NAP

Patients

Seq Age Sex Outcome Treatment
1 11 YR