FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4122476 · Received September 26, 2014

Report

Report Number
2531779-2014-27665
Event Type
Malfunction
Date Received
September 26, 2014
Report Date
September 12, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/03/2014 WITH THE FOLLOWING FINDINGS: THERE WAS NO DAMAGE FOUND IN OR AROUND THE CARTRIDGE COMPARTMENT; THE COMPLAINT COULD NOT BE CONFIRMED ON INVESTIGATION. UNRELATED TO THE ORIGINAL COMPLAINT, INVESTIGATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED IN MULTIPLE AREAS. ADDITIONALLY, THE DISPLAY SCREEN WAS DIM AND FADED AND THE DISPLAY SCREEN TEXT WAS DISCOLORED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE CARTRIDGE COMPARTMENT WAS CRACKED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601297 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1