FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4122414 · Received September 18, 2014

Report

Report Number
2032227-2014-24564
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
August 19, 2014
Report Date
August 19, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT AS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP HAS CRACKED CASE AT DISPLAY WINDOW CORNERS, BELT CLIP SLOT AND RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UP ARROW BUTTON ON THE INSULIN PUMP WAS NOT RESPONDING PROPERLY. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 2.8 MMOL/L. NO FURTHER INFORMATION REPORTED.